Numéro
J. Phys. Phys. Appl.
Volume 25, Numéro S3, mars 1964
Page(s) 23 - 30
DOI https://doi.org/10.1051/jphysap:0196400250302300
J. Phys. Phys. Appl. 25, 23-30 (1964)
DOI: 10.1051/jphysap:0196400250302300

Dispositif de mesure de viscosà 500 °C et sous pression jusqu'à 20 kg/cm 2

R. Bessouat et S. Elberg

Section des Transferts Thermiques, Centre d'Études Nucléaires de Grenoble


Abstract
The apparatus described here was built for the study of organic coolants for nuclear reactors. It uses Lawaczek's falling cylinder viscometer principle and allows measurements up to 500 °C and 20 atmospheres. A magnetic device is used to measure the falling cylinder velocity. The accuracy of viscosity determination is about ± 0.7 % at room temperature and ± 1.2 % at 500 °C. Results of measurements made with diphenyl, terphenyls OMP and OM2 and with coolants of petroleum origin are given.


Résumé
Ce dispositif a été construit pour l'étude des fluides caloporteurs organiques utilisables dans les réacteurs atomiques. Il utilise le principe du viscosimètre à chute du cylindre de Lawaczek et permet des mesures jusqu'à des températures de 500 °C et sous pression jusqu'à 20 kg/cm2. Une détection magnétique est utilisée pour la mesure de la vitesse de chute. La détermination des viscosités se fait avec une incertitude de l'ordre de ± 0,7 % à la température ambiante et de ± 1,2 % à 500 °C. Les résultats des essais effectués sur le diphényle, les terphényles OMP et OM2 et sur des fluides caloporteurs d'origine pétrolière sont fournis.

PACS
8385J - Viscosity measurements.

Key words
fission reactor materials -- viscometers