La détection des neutrons rapides à l'aide de chambres d'ionisation remplies d'hexane liquide

Daniel Blanc, Jacques Mathieu et Paul Vermande

Centre de Physique Nucléaire, Faculté des Sciences, Toulouse


Abstract
The use of a vacuum-tight polyethylene bag filled with hexane and inserted between the electrodes of the chamber will keep the liquid away from the walls and enable to obtain a residual current lower than 5 × 10-14 A/cm2, and stable as a function of time. The ionization current is a linear function of the flux, with neutrons of 14.6 MeV and 3 MeV, produced by D-T and D-D reactions ; 85 per cent of this current is attribuable to neutrons and 15 per cent to gamma photons accompanying them. The sensitivity is 100 n cm-2 s-1 (3 MeV), 1 000 n cm-2 s-1 (14.6 MeV), 10 000 photons cm-2 s-1 (60Co).


Résumé
L'emploi d'un sachet étanche de polyéthylène rempli d'hexane et serré entre les électrodes de la chambre isole le liquide des parois et permet d'obtenir un courant résiduel inférieur à 5 × 10-14 A/cm2, stable en fonction du temps. Pour des neutrons de 14,6 MeV et de 3 MeV, produits par les réactions D-T et D-D, le courant d'ionisation est une fonction linéaire du flux ; 85 % de ce courant est attribuable aux neutrons, 15 % aux photons gamma qui les accompagnent. La sensibilité est de 100 n cm-2 s-1 (3 MeV), 1 000 n cm-2 s-1 (14,6 MeV), 10 000 photons cm-2 s-1 ( 60 Co).