Numéro
J. Phys. Phys. Appl.
Volume 22, Numéro S6, juin 1961
Page(s) 121 - 132
DOI https://doi.org/10.1051/jphysap:01961002206012100
J. Phys. Phys. Appl. 22, 121-132 (1961)
DOI: 10.1051/jphysap:01961002206012100

Une installation de mesure des aimantations entre 2 et 1200 °K

Guy Rimet

Laboratoire d'Électrostatique et de Physique du Métal, Grenoble


Abstract
An apparatus for the measurement of magnetization in fields up to 30 000 oersteds at any temperature between 2 and 1 200 °K, which has been built at the Laboratoire d'Électrostatique et de Physique du Métal de Grenoble, is described in detail. The magnetization is measured with an accuracy of about 0.5 % by axial extraction out of the gap of a 150 kW electromagnet. The required temperature is obtained in an oven or in a metal-walled Dewar vessel, with an accuracy of about 0.5°. Samples up to 13 mm length and for 4 to 8 mm diameter, depending on the temperature, can be used.


Résumé
Nous décrivons en détail une installation de mesure des aimantations, construite et mise au point au Laboratoire d'Électrostatique et de Physique du métal de Grenoble, dans des champs pouvant varier jusqu'à 30 000 oersteds et à toutes les températures comprises entre 2 et 1 200 °K. Les aimantations sont mesurées avec une précision de l'ordre de 0,5 % par la méthode d'extraction axiale à partir du centre de l'entrefer d'un électro-aimant de 150 kW. Les mises en température sont effectuées à l'intérieur d'un vase isolant métallique ou d'un four à 0,5° près. On peut utiliser des échantillons ayant jusqu'à 13 mm de long et 4 à 8 mm de diamètre, selon les températures.

PACS
7560E - Magnetization curves, hysteresis, Barkhausen and related effects.
0755 - Magnetic instruments and components.

Key words
magnetic field measurement