Sur la pente du plateau des compteurs de geiger à argon et alcool éthylique

Jacques Franeau1, Fernand Grard2 et René Libert3

1  Faculté Polytechnique de Mons
2  Institut Interuniversitaire des Sciences Nucléaires (Centre annexé à la Faculté Polytechnique de Mons)
3  A. I. Ms


Abstract
The experimental study of the plateau of Geiger counters, filled with a mixture of argon and ethyl alcohol, has shown that the slope remains practically constant beyond a certain partial pressure of argon. Study of the causes of this slope made clear that at the beginning of the plateau, the cause resides in the differences of the height of the pulses ; further on, the secondary pulses become an important cause ; however these are not sufficient to explain the whole slope.


Résumé
L'étude expérimentale des facteurs influençant la pente du plateau des compteurs de Geiger, contenant un mélange d'argon et d'alcool éthylique, a montré qu'au delà d'une certaine pression partielle d'argon la pente restait pratiquement constante. L'étude des causes de cette pente a mis en évidence qu'au début du plateau, celle-ci est due à la hauteur inégale des impulsions ; ensuite ce sont les impulsions parasites qui constituent une cause importante, toutefois celles-ci ne sont pas suffisantes pour expliquer la totalité de la pente.

PACS
2940 - Radiation detectors.

Key words
Geiger counters